ECG Simulator

Phantom 320, the standard of ECG Simulators.
Most test laboratories and inspection authorities are using the handy device. The phantom 320 is the indispensable tool for quality assurance of medical engineering products, for service and test technicians or for sales demonstrations.
The friendly user interface and its compatibility to all device manufacturers, its versatileness complete the power of the phantom 320. The simulator is not only verifying the precise function of ECG and prolonged ECG devices but also checks Monitors, ECG Recorders, Arrhythmia Computers etc.
Normal or pathological ECGs: you reveive independent and genuine ECG waves for up to 32 ECG signals.

ph320e_O


Among other features you can simulate

  • 8 normal sinus rhythm
  • 7 supraventricular and
  • 9 ventricular arrhythmia
  • 3 pacemakers

A special automatic program executes a typical selection of normal and pathological signals within approx. 17 minutes.

Specifications

  • 12 leads Standard i, II ,III, avR, avL, avF, V1, V2, V3, V4, V5, V6
  • 14 connections 4 mm banana plugs and pushbuttons
  • Identification and colour code EN 60601-2-51
  • Digital storage 1,6 Mbit
  • Amplitude resolution 8 Bit
  • Scanning frequency 256 Hz
  • Signal bandwidth 0 – 120 Hz
  • Signal amplitudes
  • (min – max) Lead I + 1.35 – 1.49 mV
  • Lead II + 2.35 – 2.49 mV
  • Lead III + 0.95 – 1.06 mV
  • Lead V1 – 2.12 – 2.30 mV
  • Lead V2 – 0.55 – 0.61 mV
  • Lead V3 + 1.03 – 1.14 mV
  • Lead V4 + 2.11 – 2.21 mV
  • Lead V5 + 1.75 – 1.85 mV
  • Lead V6 + 1.40 – 1.51 mV
  • Time base Quarz 32768 Hz
  • Output impedance Electrode connections: 20 Ohm
  • 1V outputs: 100 Ohm
  • Waveforms 8 sinus rhythms (normal QRS)
  • 30,45,60,75,90,120,150,180 BPM
  • 7 supraventricular arrhythmia
  • 9 ventricular arrhythmia
  • 3 pacemaker
  • 2 ST segments (elevation, depression)
  • Each with horizontal, ascending and descending ST wave
  • Bradycardia
  • Tachycardia
  • ECG with artefacts
  • ECG with interference 50/60 Hz
  • Signal width PQ, QRS, QT depending on RR interval (after Lepeschkin)
  • Battery Standard 2 x 1.5-V LR 6
  • Battery life Approx. 100 hours continuous operation
  • Surroundings Working temperature 0 bis 40 °C
  • Storage temperature -20 bis 80 °C
  • Air humidity: 10% – 90% noncondensing
  • Barometric pressure 700 – 1500 hPa
  • Dimensions (LxWxH) 150 x 100 x 40 mm
  • Weight 400 grams, including battery

Download the data sheet